Technical specifications WSU 2 plus
• Economical and resource-saving production of ultrapure water for laboratory use and in-vitro diagnostics
• Simple and quick (< 1 min.) consumables replacement by the user
• Simple and safe sampling point
Operating principle
Multi-stage ion exchanger with UV disinfection reverse osmosis module and sterile filter units
Up to 10 times longer resin service life compared to conventional water treatment systems without a reverse osmosis module.
Output Conductivity
TOC-value (Total Organic Carbon)
Microbiological retention
<1.0 µS/cm
<500 ppb
<1 cfu/ml
Highlights
• Web-based interface for real-time data monitoring
• Mobile frame, trolley
• Recommended: Pre-filtration unit
- Pre-filtration unit for decent water quality, mountable on trolley
- Pre-filtration unit for pure and very pure water quality, on trolley
• Leakage sensor included
• WSU 2 plus: 70 L/h or 120 L/h withdrawal volume,
• WSU 2 mini: 20 L/h withdrawal volume
• Meets the Specification of all cobas instruments by Roche Diagnostics
• Temperature-compensated conductivity measurement
• Dynamic maintenance intervals depending on consumption and water input quality
(The device displays when something needs to be changed)
• Independent emergency supply function with conductivity monitoring
• 10 L reservoir tank included (up to 10 L of ultrapure water can be extracted, after failure of water or power supply)
• Quick exchange of consumables (resin, polisher, filters, UV-lamp) < 1 min.
• 10 times longer resin and polisher lifetime due to reverse osmosis membrane (compared to conventional systems)
-
Control unit:
- Microcontroller system with touch display
- Fixed alarm for resistivity 1µS/cm
- Network connection for remote monitoring
-
Sterile filter:
- 2x sterile filter with filtrate connections
- Membrane: 0,2 µm
- Filtration performance: 2,5 L/min
- Max. filtration pressure: 2,0 bar
-
Pre-filtration Unit:
- 2x sterile filter with filtrate connections
- Membrane: 0,2 µm
- Filtration performance: 2,5 L/min
- Max. filtration pressure: 2,0 bar
-
UV-lamp:
- UV-lamp 14W, lifetime approx. 2000-3000 hours
-
Spectrum Pure Cartridge:
- The Spectrum Pure Cartridge is specifically designed for the WSU 2 plus device range to produce ultrapure water with a conductivity >18MOhm.
- The Spectrum Pure Cartridge is ready to use filled with a strongly acidic cation exchanger and a strong alkaline anion exchanging resin.
-
Spectrum Ultra Pure Cartridge:
- The Spectrum Pure Cartridge is specifically designed for the WSU 2 plus device range to produce ultrapure water with a conductivity >18MOhm.
- The Spectrum Pure Cartridge is ready to use filled with a strongly acidic cation exchanger and a strong alkaline anion exchanging resin. Both have a monodisperse particle size distribution based on a styrene divinylbenzene copolymer.
All water-wetted parts are made out of plastic or stainless steel, so that the WSU 2 plus can be operated with tap water as well as with deionized water.
WSU 2 plus Order Information
Roche Material Number | Description | |
---|---|---|
09612351001 | Water Supply Unit 2 plus, 70 L/h, with Trolley | |
09619763001 | Water Supply Unit 2 plus, 120 L/h, with Trolley | |
Consumables | ||
09806415001 | RO-membrane WSU 2 plus, 70 L/h | |
09808531001 | RO-membrane WSU 2 plus, 120 L/h | |
08452296001 | UV-lamp WSU 2 plus | |
09870504001 | Spectrum Pure Cartridge filled | |
09870512001 | Spectrum Ultra Pure Cartridge filled | |
08452261001 | Sterile filter WSU 2 plus (0,2µm) | |
08452300001 | Filter element active coal 4 7/8″ (0.5µm) | |
Spare part | ||
09806431001 | Pressure vessel for RO-membrane |